Low Sensitivity in HPLC and UHPLC: Causes, Diagnostics, and Corrective Actions
A comprehensive technical guide to identifying and resolving sensitivity issues in high-performance liquid chromatography systems
Understanding the Challenge
High-performance liquid chromatography (HPLC) and ultra-high-performance liquid chromatography (UHPLC) are widely used analytical techniques for trace-level qualitative and quantitative analysis. One of the most frequently reported performance issues in both systems is low sensitivity, characterized by reduced peak height, poor signal-to-noise ratio, and elevated detection limits.
Low sensitivity directly impacts method robustness, quantitative accuracy, and regulatory compliance. Identifying the underlying cause requires a systematic evaluation of sample introduction, detector configuration, extracolumn effects, system integrity, and mobile-phase conditions.
Symptom and Observable Problem
Low sensitivity in HPLC or UHPLC may present as one or more of the following:
Reduced Response
Reduced peak height or peak area compared to historical performance
Peak Distortion
Broad, distorted, or flattened chromatographic peaks
Early Elution
Early analyte elution with poor resolution
Baseline Noise
Increased baseline noise relative to analyte response
Poor Reproducibility
Poor reproducibility of peak response between injections
These symptoms often occur even when system pressure and flow appear nominal, making low sensitivity a diagnostic challenge without a structured approach.
Root Cause Analysis
Low sensitivity is typically caused by analyte dilution, peak broadening, or signal loss occurring either at injection, during separation, or at detection. The most common contributors are outlined below.
1. Sample Overload Leading to Early Analyte Elution
Excessive sample concentration or injection volume can exceed the stationary phase loading capacity. When this occurs, analytes elute prematurely with broadened and distorted peaks due to incomplete retention. The result is a lower peak height and reduced detector response despite higher analyte mass.
2. Detector Cell Volume Too Large
Large detector flow cells increase band broadening within the detection zone. This internal dispersion reduces peak height and dilutes the analyte signal, directly decreasing sensitivity. This effect is especially pronounced in UHPLC systems where peak volumes are inherently small.
3. Injection Volume Too Large or Incompatible Injection Solvent
Large injection volumes or the use of a strong injection solvent relative to the mobile phase prevents proper analyte focusing at the column inlet. Poor focusing results in broader peaks and reduced signal intensity, even when analyte mass is sufficient.
4. Excessive Extracolumn Volume (Dead Volume)
Extracolumn volume from tubing, fittings, connectors, and detector cells causes additional band broadening outside the column. In UHPLC systems, where column volumes are small, even minor dead volume can significantly degrade sensitivity.
5. System Leaks
Leaks at fittings, injector seals, or detector connections can reduce effective analyte mass reaching the detector. Leaks may also introduce baseline noise, pressure instability, and flow inconsistencies that further reduce sensitivity.
6. Mobile Phase Viscosity and Composition
Highly viscous mobile phases increase mass-transfer resistance and can broaden peaks. Inappropriate solvent strength may lead to insufficient retention, early elution, and poor peak shape, all of which reduce sensitivity.
7. Pressure Instability During Separation
Fluctuations or unexpected drops in column pressure may indicate partial blockages, leaks, or pump irregularities. These conditions disrupt flow stability and compromise peak integrity and detector response.
Diagnostic Workflow
Diagnostics
A disciplined diagnostic workflow improves root-cause identification and minimizes unnecessary component replacement.
Evaluate chromatograms
for peak broadening, early elution, and signal-to-noise degradation
Review baseline stability
for noise or drift indicative of leaks or flow irregularities
Trend system pressure
for sudden drops, instability, or abnormal behavior
Confirm injection parameters
including volume and injection solvent strength
Inspect detector flow cell specifications
and cleanliness
Estimate extracolumn volume
by reviewing tubing length, internal diameter, and fitting geometry
Perform a dilution test
by diluting the sample (e.g., 1:10) and reinjecting to assess overload effects
Review mobile phase composition
and temperature settings
Solutions
Corrective Actions
Corrective actions should be implemented incrementally and verified by reinjection after each change.
Sample Overload Mitigation
Dilute the sample (commonly 1:10) and reinject
Reduce injection volume
Use a column with higher loading capacity if required
Detector Optimization
Select the smallest practical detector cell volume compatible with the method
Verify flow cell cleanliness and proper installation
Injection Optimization
Reduce injection volume to less than one-sixth of the volume of the first-eluting peak
Match or weaken injection solvent strength relative to the mobile phase to improve focusing
Additional Corrective Measures
Extracolumn Volume Reduction
Replace fittings with low- or zero-dead-volume connectors
Use the shortest tubing lengths possible
Select appropriate internal diameters (<0.010 inch for HPLC; <0.005 inch for UHPLC)
Ensure proper fitting seating and alignment
Leak Elimination and Pressure Stabilization
Reseat or replace leaking fittings
Replace worn injector seals or pump components as needed
Remove blockages and confirm unrestricted flow through the system
Mobile Phase and Temperature Optimization
Increase column temperature to reduce solvent viscosity and improve mass transfer
Adjust mobile phase composition to improve retention without excessive viscosity
Related Issues
Low sensitivity is often associated with or may contribute to additional chromatographic problems, including:
Poor quantitative accuracy and precision
Method robustness failures during validation
Increased variability between injections
Shortened column lifetime due to overload conditions
Misinterpretation of detection limits and reporting thresholds